Determination of Embryo/Fetus Doses of Pregnants in Thorax Computed Tomography Scan
TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, cilt.38, ss.476-481, 2023 (ESCI, Scopus, TRDizin)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 38
- Basım Tarihi: 2023
- Doi Numarası: 10.5505/tjo.2023.4066
- Dergi Adı: TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY
- Derginin Tarandığı İndeksler: Emerging Sources Citation Index (ESCI), Scopus, Academic Search Premier, CINAHL, TR DİZİN (ULAKBİM)
- Sayfa Sayıları: ss.476-481
- Açık Arşiv Koleksiyonu: AVESİS Açık Erişim Koleksiyonu
- İstanbul Üniversitesi-Cerrahpaşa Adresli: Evet
Özet
OBJECTIVE The aim of this retrospective study was to experimentally estimate the embryo/fetal dose from non conformal collimation computed tomography (CT) scans of pregnant women during thorax computed tomography (CCT) scans. METHODS The study involved 22 pregnant women who underwent CCT scans unintentionally during 2015-2022. Embryo/fetal absorbed doses were calculated by FetDose V4 software and compared with experimental measures using Alderson Rando Phantom and thermoluminescent dosimeters (TLDs). RESULTS The CCT of 6 patients (27%) showed the thorax region in the scanned area only; whereas 16 patients (73%) exhibited a wider scan length involving the thorax and abdomen. The average embryo/fetal dose was found to be 1.54 & PLUSMN;0.71 mGy for 6 patients who had only the thorax area in the CCT, whereas the mean embryo/fetal dose of the 16 patients with thorax and abdomen CCT was calculated as 5.31 & PLUSMN;1.67 mGy. CONCLUSION As a result, non-conformal collimation was observed in the abdomen and thorax in CCT scans leading to a relative increase in the embryo/fetal dose.